oai:arXiv.org:1404.6303 [pdf] - 814851
A linear calibration method on DNL error for energy spectrum
Bo, Lu;
Yan-Hong, Fu;
Yong, Chen;
Wei-Wei, Cui;
Yu-Sa, Wang;
Yu-Peng, Xu;
Guo-He, Yin;
Yan-Ji, Yang;
Yi, Zhang;
Jia, Huo;
Wei, Li;
Tian-Xiang, Cheng;
Wei, Hu;
Da-Wei, Han;
Mao-Shun, Li;
Juan, Wang;
Yu, Wang;
Xiao-Yang, Liu;
Yue, Zhu;
Zhong-Yi, Zhao;
Ya, Zhang
Submitted: 2014-04-24
A calibration method aimed for the differential nonlinearity (DNL) error of
the Low Energy X-ray Instrument (LE) onboard the Hard X-ray Modulation
Telescope (HXMT) is presented, which is independent with electronic systems
used as testing platform and is only determined by the analog-to-digital
converter (ADC) itself. Exploring this method, ADCs that are used within the
flight model phase of HXMT-LE can be calibrated individually and independently
by a non-destructive and low-cost way, greatly alleviating the complexity of
the problem. As a result, the performance of the energy spectrum can be
significantly improved, further more, noise reduced and resolution enhanced.
[2]
oai:arXiv.org:1308.1698 [pdf] - 704779
Measurements of Charge Transfer Efficiency in a Proton-irradiated Swept
Charge Device
YuSa, Wang;
Yong, Chen;
XiaoYan, Liu;
WeiWei, Cui;
YuPeng, Xu;
ChengKui, Li;
MaoShun, Li;
DaWei, Han;
TianXiang, Chen;
Jia, Huo;
Juan, Wang;
Wei, Li;
Wei, Hu;
Yi, Zhang;
Bo, Lu;
GuoHe, Yin;
Yue, Zhu;
ZiLiang, Zhang
Submitted: 2013-08-07
Charge Coupled Devices (CCDs) have been successfully used in several low
energy X-ray astronomical satellite over the past two decades. Their high
energy resolution and high spatial resolution make them an perfect tool for low
energy astronomy, such as formation of galaxy clusters and environment of black
holes. The Low Energy X-ray Telescope (LE) group is developing Swept Charge
Device (SCD) for the Hard X-ray Modulation Telescope (HXMT) satellite. SCD is a
special low energy X-ray CCD, which could be read out a thousand times faster
than traditional CCDs, simultaneously keeping excellent energy resolution. A
test method for measuring the charge transfer efficiency (CTE) of a prototype
SCD has been set up. Studies of the charge transfer inefficiency (CTI) have
been performed at a temperature range of operation, with a proton-irradiated
SCD.