[1]  oai:arXiv.org:astro-ph/0605500  [pdf] - 82187
The ZEPLIN-III dark matter detector: instrument design, manufacture and commissioning
Comments: 25 pages, 19 figures. Submitted to Astropart. Phys. Some figures down sampled to reduce size
Submitted: 2006-05-19
We present details of the technical design and manufacture of the ZEPLIN-III dark matter experiment. ZEPLIN-III is a two-phase xenon detector which measures both the scintillation light and the ionisation charge generated in the liquid by interacting particles and radiation. The instrument design is driven by both the physics requirements and by the technology requirements surrounding the use of liquid xenon. These include considerations of key performance parameters, such as the efficiency of scintillation light collection, restrictions placed on the use of materials to control the inherent radioactivity levels, attainment of high vacuum levels and chemical contamination control. The successful solution has involved a number of novel design and manufacturing features which will be of specific use to future generations of direct dark matter search experiments as they struggle with similar and progressively more demanding requirements.