Normalized to: Yorita, K.
[1]
oai:arXiv.org:2003.14248 [pdf] - 2077149
Liquid argon scintillation response to electronic recoils between
$2.8$--$1275~{\rm keV}$ in a high light yield single-phase detector
Submitted: 2020-03-31, last modified: 2020-04-10
We measure the liquid argon scintillation response to electronic recoils in
the energy range of $2.82$ to $1274.6~{\rm keV}$. The single-phase detector
with a large optical coverage used in this measurement yields $12.6 \pm 0.3 ~
(11.1 \pm 0.3)~{\rm photoelectron/keV}$ for $661.7~{\rm keV}$ $\gamma$-ray
events based on a photomultiplier tube single photoelectron response modeling
with a Gaussian plus an additional exponential terms (with only Gaussian term).
It is exposed to a variety of calibration sources such as $^{137}{\rm Cs}$ and
$^{241}{\rm Am}$ $\gamma$-ray emitters, and $^{252}{\rm Cf}$ fast neutron
emitter that induces quasimonoenergetic $\gamma$-rays through a $(n, n'\gamma)$
reaction with $^{19}{\rm F}$ in polytetrafluoroethylene. In addition, the high
light yield enables identification of the $2.82~{\rm keV}$ peak of $^{37}{\rm
Ar}$, a cosmogenic isotope in atmospheric argon. The scintillation yield and
energy resolution of the detector are obtained by the full-absorption peaks. We
find up to approximately $25\%$ shift in the scintillation efficiency across
the energy range and less than $3\%$ of the energy resolution for the
$661.7~{\rm keV}$ line. The energy dependent scintillation quenching can be
attributed by the electron-ion recombination process, and is discussed by an
analogy to the dependence of liquid xenon. The Thomas-Imel Box model with its
constant parameter $\varsigma=0.07 ^{+0.03}_{-0.02}$ is found to explain the
results below $200~{\rm keV}$.