oai:arXiv.org:1103.3689 [pdf] - 966518
Characterization of the QUartz Photon Intensifying Detector (QUPID) for
Noble Liquid Detectors
Teymourian, A.;
Aharoni, D.;
Baudis, L.;
Beltrame, P.;
Brown, E.;
Cline, D.;
Ferella, A. D.;
Fukasawa, A.;
Lam, C. W.;
Lim, T.;
Lung, K.;
Meng, Y.;
Muramatsu, S.;
Pantic, E.;
Suyama, M.;
Wang, H.;
Arisaka, K.
Submitted: 2011-03-18, last modified: 2011-07-15
Dark Matter and Double Beta Decay experiments require extremely low
radioactivity within the detector materials. For this purpose, the University
of California, Los Angeles and Hamamatsu Photonics have developed the QUartz
Photon Intensifying Detector (QUPID), an ultra-low background photodetector
based on the Hybrid Avalanche Photo Diode (HAPD) and entirely made of
ultraclean synthetic fused silica. In this work we present the basic concept of
the QUPID and the testing measurements on QUPIDs from the first production
line. Screening of radioactivity at the Gator facility in the Laboratori
Nazionali del Gran Sasso has shown that the QUPIDs safely fulfill the low
radioactive contamination requirements for the next generation zero background
experiments set by Monte Carlo simulations. The quantum efficiency of the QUPID
at room temperature is > 30% at the xenon scintillation wavelength. At low
temperatures, the QUPID shows a leakage current less than 1 nA and a global
gain of 10^5. In these conditions, the photocathode and the anode show > 95%
linearity up to 1 uA for the cathode and 3 mA for the anode. The photocathode
and collection efficiency are uniform to 80% over the entire surface. In
parallel with single photon counting capabilities, the QUPIDs have a good
timing response: 1.8 +/- 0.1 ns rise time, 2.5 +/- 0.2 ns fall time, 4.20 +/-
0.05 ns pulse width, and 160 +/- 30 ps transit time spread. The QUPIDs have
also been tested in a liquid xenon environment, and scintillation light from
57Co and 210Po radioactive sources were observed.