Normalized to: Stork, M.
[1]
oai:arXiv.org:1412.6457 [pdf] - 911181
Design and Performance of the X-ray polarimeter X-Calibur
Beilicke, Matthias;
Kislat, F.;
Zajczyk, A.;
Guo, Q.;
Endsley, R.;
Stork, M.;
Cowsik, R.;
Dowkontt, P.;
Barthelmy, S.;
Hams, T.;
Okajima, T.;
Sasaki, M.;
Zeiger, B.;
De Geronimo, G.;
Baring, M. G.;
Krawczynski, H.
Submitted: 2014-12-19
X-ray polarimetry promises to give qualitatively new information about
high-energy astrophysical sources, such as binary black hole systems,
micro-quasars, active galactic nuclei, neutron stars, and gamma-ray bursts. We
designed, built and tested a X-ray polarimeter, X-Calibur, to be used in the
focal plane of the balloon-borne InFOCuS grazing incidence X-ray telescope.
X-Calibur combines a low-Z scatterer with a CZT detector assembly to measure
the polarization of 20-80keV X-rays making use of the fact that polarized
photons scatter preferentially perpendicular to the electric field orientation.
X-Calibur achieves a high detection efficiency of ~80%. The X-Calibur detector
assembly is completed, tested, and fully calibrated. The response to a
polarized X-ray beam was measured successfully at the Cornell High Energy
Synchrotron Source. This paper describes the design, calibration and
performance of the X-Calibur polarimeter. In principle, a similar space-borne
scattering polarimeter could operate over the broader 2-100keV energy band.