oai:arXiv.org:1506.08212 [pdf] - 1269370
The DICE calibration project: design, characterization, and first
results
Regnault, N.;
Guyonnet, A.;
Schahmanèche, K.;
Guillou, L. Le;
Antilogus, P.;
Astier, P.;
Barrelet, E.;
Betoule, M.;
Bongard, S.;
Cuillandre, J. -C.;
Juramy, C.;
Pain, R.;
Rocci, P. -F.;
Tisserand, P.;
Villa, F.
Submitted: 2015-06-26
We describe the design, operation, and first results of a photometric
calibration project, called DICE (Direct Illumination Calibration Experiment),
aiming at achieving precise instrumental calibration of optical telescopes. The
heart of DICE is an illumination device composed of 24 narrow-spectrum,
high-intensity, light-emitting diodes (LED) chosen to cover the
ultraviolet-to-near-infrared spectral range. It implements a point-like source
placed at a finite distance from the telescope entrance pupil, yielding a flat
field illumination that covers the entire field of view of the imager. The
purpose of this system is to perform a lightweight routine monitoring of the
imager passbands with a precision better than 5 per-mil on the relative
passband normalisations and about 3{\AA} on the filter cutoff positions. The
light source is calibrated on a spectrophotometric bench. As our fundamental
metrology standard, we use a photodiode calibrated at NIST. The radiant
intensity of each beam is mapped, and spectra are measured for each LED. All
measurements are conducted at temperatures ranging from 0{\deg}C to 25{\deg}C
in order to study the temperature dependence of the system. The photometric and
spectroscopic measurements are combined into a model that predicts the spectral
intensity of the source as a function of temperature. We find that the
calibration beams are stable at the $10^{-4}$ level -- after taking the slight
temperature dependence of the LED emission properties into account. We show
that the spectral intensity of the source can be characterised with a precision
of 3{\AA} in wavelength. In flux, we reach an accuracy of about 0.2-0.5%
depending on how we understand the off-diagonal terms of the error budget
affecting the calibration of the NIST photodiode. With a routine 60-mn
calibration program, the apparatus is able to constrain the passbands at the
targeted precision levels.
[2]
oai:arXiv.org:1208.6301 [pdf] - 557215
(Sn)DICE: A Calibration System Designed for Wide Field Imagers
Submitted: 2012-08-30
Dark Energy studies with type Ia supernovae set very tight constraints on the
photometric calibration of the imagers used to detect the supernovae and follow
up their flux variations. Among the key challenges is the measurement of the
shape and normalization of the instrumental throughput. The DICE system was
developed by members of the Supernova Legacy Survey (SNLS), building upon the
lessons learnt working with the MegaCam imager. It consists in a very stable
light source, placed in the telescope enclosure, and generating compact,
conical beams, yielding an almost flat illumination of the imager focal plane.
The calibration light is generated by narrow spectrum LEDs selected to cover
the entire wavelength range of the imager. It is monitored in real time using
control photodiodes. A first DICE demonstrator, SnDICE has been installed at
CFHT. A second generation instrument (SkyDICE) has been installed in the
enclosure of the SkyMapper telescope. We present the main goals of the project.
We discuss the main difficulties encountered when trying to calibrate a wide
field imager, such as MegaCam (or SkyMapper) using such a calibrated light
source.