oai:arXiv.org:1807.04158 [pdf] - 1714414
3D metrology with a laser tracker inside a vacuum chamber for NISP test
campaign
Submitted: 2018-07-11
In the frame of the test of NISP instrument for ESA Euclid mission, the
question was raised to perform a metrology measurement of different components
during the thermal vacuum test of NISP instrument. NISP will be tested at
Laboratoire d'Astrophysique de Marseille (LAM) in ERIOS chamber under vacuum
and thermal conditions in order to qualify the instrument in its operating
environment and to perform the final acceptance test before delivery to the
payload. One of the main objectives of the test campaign will be the
measurement of the focus position of NISP image plane with respect to the
EUCLID object plane. To simulate the EUCLID object plane, a telescope simulator
with a very well know focal distance will be installed in front of NISP into
ERIOS chamber. We need to measure at cold and vacuum the position of reflectors
installed on NISP instrument and the telescope simulator. From these
measurements, we will provide at operational temperature the measurement of
references frames set on the telescope simulator and NISP, the knowledge of the
coordinates of the object point source provided by the telescope simulator and
the measurement of the angle between the telescope simulator optical axis and
NISP optical axis. In this context, we have developed a metrology method based
on the use of a laser tracker to measure the position of the reflectors inside
ERIOS. The laser tracker is installed outside the vacuum chamber and measure
through a curved window reflectors put inside the chamber either at ambient
pressure or vacuum pressure. Several tests campaigns have been done at LAM to
demonstrate the measurement performance with this configuration. Using a well
know reflectors configuration, we show that it is possible to correct the laser
tracker measurement from the window disturbances and from the vacuum impact. A
corrective term is applied to the data and allows retrieving the real
coordinates of the reflectors with a bias lower than 30$\mu$m, which is lower
than the laser tracker measurement uncertainties estimated at 60$\mu$m. No
additional error term of the laser tracker measurement is observed when using
the laser tracker with the curved window and in vacuum, comparing with a
classical use of the laser tracker. With these test campaign, we have been able
to demonstrate the possibility to use a laser tracker to measure in real time
during a vacuum thermal test the position of different mechanical parts into a
vacuum chamber with an accuracy better than 60$\mu$m.