Performance of a 4 Kelvin pulse-tube cooled cryostat with dc SQUID
amplifiers for bolometric detector testing
Submitted: 2013-01-04
The latest generation of cosmic microwave background (CMB) telescopes is
searching for the undetected faint signature of gravitational waves from
inflation in the polarized signal of the CMB. To achieve the unprecedented
levels of sensitivity required, these experiments use arrays of superconducting
Transition Edge Sensor (TES) bolometers that are cooled to sub-Kelvin
temperatures for photon-noise limited performance. These TES detectors are read
out using low- noise SQUID amplifiers. To rapidly test these detectors and
similar devices in a laboratory setting, we constructed a cryogenic
refrigeration chain consisting of a commercial two-stage pulse-tube cooler,
with a base temperature of 3 K, and a closed-cycle 3He/4He/3He sorption cooler,
with a base temperature of 220 mK. A commercial dc SQUID system, with sensors
cooled to 4 K, was used as a highly-sensitive cryogenic ammeter. Due to the
extreme sensitivity of SQUIDs to changing magnetic fields, there are several
challenges involving cooling them with pulse-tube coolers. Here we describe the
successful design and implementation of measures to reduce the vibration,
electromagnetic interference, and other potential sources of noise associated
with pulse-tube coolers.