Perepelitsyn, Alexander E.
Normalized to: Perepelitsyn, A.
5 article(s) in total. 9 co-authors, from 1 to 3 common article(s). Median position in authors list is 4,0.
[1]
oai:arXiv.org:2005.14598 [pdf] - 2104109
Mapper of Narrow Galaxy Lines (MaNGaL): new tunable filter imager for
Caucasian telescopes
Submitted: 2020-05-29
We described the design and operation principles of a new tunable-filter
photometer developed for the 1-m telescope of the Special Astrophysical
Observatory of the Russian Academy of Sciences and the 2.5-m telescope of the
Sternberg Astronomical Institute of the Moscow State University. The instrument
is mounted on the scanning Fabry-Perot interferometer operating in the
tunable-filter mode in the spectral range of 460-800 nm with a typical spectral
resolution of about 1.5 nm. It allows one to create images of galactic and
extragalactic nebulae in the emission lines having different excitation
conditions and to carry out diagnostics of the gas ionization state. The main
steps of observations, data calibration, and reduction are illustrated by
examples of different emission-line objects: galactic HII regions, planetary
nebulae, active galaxies with extended filaments, starburst galaxies, and
Perseus galaxy cluster.
[2]
oai:arXiv.org:1911.10984 [pdf] - 2038350
Medium-band photometric reverberation mapping of AGNs at $0.1 < z <
0.8$. Techniques and sample
Submitted: 2019-11-21
The most popular method of the broad-line region size estimation in active
galactic nuclei (AGN) is the reverberation mapping based on measuring the time
delay between the continuum flux and the flux in the emission lines. In our
work, we apply the method of photometric reverberation mapping in mid-band
filters, adapted for observations on the 1-m Zeiss-1000 telescope of Special
Astrophysical Observatory of Russian Academy of Sciences, for the study of AGN
with broad lines in the range of redshifts $0.1 < z < 0.8$. This paper provides
a sample of 8 objects, describes the technique of observations and data
processing for 2 studied objects to demonstrate the stability of the used
method.
[3]
oai:arXiv.org:1911.04073 [pdf] - 2026294
Photometric reverberation mapping of AGNs at 0.1 < z < 0.8: I.
Observational technique
Submitted: 2019-11-10
The improvement of the calibration relation for determining the size of the
broad-line region from the observed optical luminosity of active galactic
nuclei (AGN) is a necessary task to study fundamental parameters of distant
AGNs such as mass of the central supermassive black hole. The most popular
method of the BLR size estimation is the reverberation mapping based on
measuring the time delay between the continuum flux and the flux in the
emission lines. In our work, we apply the method of photometric reverberation
mapping in medium-band filters, adapted for observations at the Zeiss-1000
telescope of the SAO RAS, for the study of AGN with broad lines in the range of
redshifts $0.1 < z < 0.8$. This paper describes the technique of observations
and data processing, provides a sample of objects and demonstrates the
stability of the used method. As a preliminary result for 2MASX
J08535955+7700543 at $z=0.1$ we have obtained time delay estimates of $\tau
(ICCF)=32.2^{\pm10.6}$ days and $\tau(JAVELIN)=39.5_{-15.8}^{+23.0}$ days that
are consistent with each other and also within the accuracy of the existing
calibration relations.
[4]
oai:arXiv.org:1808.09416 [pdf] - 1753290
IFU Unit in SCORPIO-2 Focal Reducer for Integral-Field Spectroscopy on
the 6-m Telescope of the SAO RAS
Submitted: 2018-08-28
We describe the scheme and design features of the new IFU unit (Integral
Field Unit) meant to perform integral-field spectroscopy as a part of SCORPIO-2
focal reducer, which is mounted in the prime focus of the 6-m telescope of the
Special Astrophysical Observatory of the Russian Academy of Sciences. The
design of the unit is based on the principle of the formation of array spectra
using a lens raster combined with optical fibers. The unit uses a rectangular
raster consisting of 22x22 square 2-mm diameter lenses. The image of the object
is transferred by an optical system with a 23x magnification from the focal
plane of the telescope to the plane of the lens raster. The image scale is
0.75"/lens and the field of view of the instrument has the size of 16.5"x16.5".
The raster also contains two extra 2x7 lens arrays to acquire the night-sky
spectra whose images are offset by $\pm$3' from the center. Optical fibers are
used to transform micropupil images into two pseudoslits located at the IFU
collimator entrance. When operating in the IFU mode a set of volume phase
holographic gratings (VPHG) provides a spectral range of 4600-7300 \AA\ and a
resolution $\lambda/\delta\lambda$ of 1040 to 2800. The quantum efficiency of
SCORPIO-2 field spectroscopy is 6-13% depending on the grating employed. We
describe the technique of data acquisition and reduction using IFU unit and
report the results of test observations of the Seyfert galaxy Mrk 78 performed
on the 6-m telescope of the Special Astrophysical Observatory of the Russian
Academy of Sciences.
[5]
oai:arXiv.org:1711.03569 [pdf] - 1602779
SCORPIO-2 Guiding and Calibration System in the Prime Focus of the 6-m
Telescope
Submitted: 2017-11-09
We describe a device (adapter) for off-axis guiding and photometric
calibration of wide-angle spectrographs operating in the prime focus of the 6-m
telescope of the Special Astrophysical Observatory of the Russian Academy of
Sciences. To compensate coma in off-axis star images an achromatic lens
corrector is used, which ensures maintaining image quality (FWHM) at a level of
about 1'' within 15' from the optical axis. The device has two 54'-diameter
movable guiding fields, which can move in 10' x 4'.5 rectangular areas. The
device can perform automatic search for guiding stars, use them to control the
variations of atmospheric transmittance, and focus the telescope during
exposure. The limiting magnitude of potential guiding stars is mR ~17 mag. The
calibration path whose optical arrangement meets the telecentrism condition
allows the spectrograph to be illuminated both by a source of line spectrum (a
He-Ne-Ar filled lamp) and by a source of continuum spectrum. The latter is
usually represented either by a halogen lamp or a set of light-emitting diodes,
which provide illumination of approximately uniform intensity over the
wavelength interval from 350 to 900 nm. The adapter is used for observations
with SCORPIO-2 multimode focal reducer.