oai:arXiv.org:0708.1924 [pdf] - 1934159
The Absolute, Relative and Multi-Wavelength Calibration of the Pierre
Auger Observatory Fluorescence Detectors
Knapik, R.;
Bauleo, P.;
Becker, B. R.;
Brack, J.;
Caruso, R.;
Fratte, C. Delle;
Dorofeev, A.;
Harton, J.;
Insolia, A.;
Matthews, J. A. J.;
Menshikov, A.;
Ortolani, F.;
Petrinca, P.;
Pichel, A.;
Riggi, S.;
Roberts, M.;
Martino, J. Rodriguez;
Rovero, A. C.;
Scuderi, M.;
Tamashiro, A.;
Torresi, D.;
Tuci, V.;
Wiencke, L.
Submitted: 2007-08-14
Absolute calibration of the Pierre Auger Observatory fluorescence detectors
uses a 375 nm light source at the telescope aperture. This end-to-end technique
accounts for the combined effects of all detector components in a single
measurement. The relative response has been measured at wavelengths of 320,
337, 355, 380 and 405 nm, defining a spectral response curve which has been
normalized to the absolute calibration. Before and after each night of data
taking a relative calibration of the phototubes is performed. This relative
calibration is used to track both short and long term changes in the detector's
response. A cross check of the calibration in some phototubes is performed
using an independent laser technique. Overall uncertainties, current results
and future plans are discussed.