Normalized to: Naebert, T.
[1]
oai:arXiv.org:1010.5711 [pdf] - 347444
Analysis of the application of the optical method to the measurements of
the water vapor content in the atmosphere - Part 1: Basic concepts of the
measurement technique
Galkin, V. D.;
Immler, F.;
Alekseeva, G. A.;
Berger, F. -H.;
Leiterer, U.;
Naebert, T.;
Nikanorova, I. N.;
Novikov, V. V.;
Pakhomov, V. P.;
Sal'nikov, I. B.
Submitted: 2010-10-27, last modified: 2011-04-20
We retrieved the total content of the atmospheric water vapor (or Integrated
Water Vapor, IWV) from extensive sets of photometric data obtained since 1995
at Lindenberg Meteorological Observatory with star and sun photometers.
Different methods of determination of the empirical parameters that are
necessary for the retrieval are discussed. The instruments were independently
calibrated using laboratory measurements made at Pulkovo Observatory with the
VKM-100 multi-pass vacuum cell. The empirical parameters were also calculated
by the simulation of the atmospheric absorption by water vapor, using the
MODRAN-4 program package for different model atmospheres. The results are
compared to those presented in the literature, obtained with different
instruments and methods of the retrieval. The reliability of the empirical
parameters, used for the power approximation that links the water vapor content
with the observed absorption, is analyzed. Currently, the total (from
measurements, calibration, and calculations) errors yield the standard
uncertainty of about 10% in the total column water vapor. We discuss the
possibilities for improving the accuracy of calibration to ~1% as indispensable
condition in order to make it possible to use data obtained by optical
photometry as an independent reference for other methods (GPS, MW-radiometers,
lidar, etc).
[2]
oai:arXiv.org:1010.3567 [pdf] - 262343
The Laboratory Complex for the Calibration of Photometers Using the
Optical Method for Determination of the Water Vapor Content in the Earth
Atmosphere
Submitted: 2010-10-18, last modified: 2010-11-15
We describe the laboratory complex for the calibration of photometers that
are used in weather service to measure the water vapor content in the Earth
atmosphere. The complex was built up in Pulkovo Observatory and developed
within the framework of collaboration between Pulkovo Observatory and
Lindenberg Meteorological Observatory (Meteorologisches Observatorium
Lindenberg - Richard-A{\ss}mann-Observatorium, Lindenberg, Germany). It is used
to obtain calibration dependences for individual devices, and also to develop
and compare various methods of construction of calibration dependences. These
techniques are based on direct calibration of the photometers, on the use of
spectral laboratory transmission functions for water vapor, on calculation
methods using spectroscopical databases for individual lines. We hope that when
the parameters of the equipment are taken into account in detail and new
results for the absorptive power of water vapor are used, the accuracy of
determination of the water vapor content in the atmosphere of 1-2% may be
attained.
[3]
oai:arXiv.org:1010.3572 [pdf] - 262344
The Determination of the Water Vapor Content in the Pulkovo VKM-100
Multipass Vacuum Cell Using Polymer Sensors of Humidity
Submitted: 2010-10-18, last modified: 2010-11-15
In spectral studies of water vapor under laboratory conditions (determination
of molecular constants, measurement for spectral transmission functions), the
amount of water vapor in the time of the measurements is one of the most
essential parameters, which should be determined accurately. We discuss the
application for this purpose of polymer sensors of humidity manufactured by
Praktik-NC (Moscow) and used in the Pulkovo VKM-100 multipass vacuum cell.
These sensors were examined in the laboratory of Lindenberg Meteorological
observatory (Germany) by comparison between their readings and those of
standard measuring devices for various values of relative humidity, pressure,
and temperature. We also carried out measurements of relative humidity in boxes
with saline solution, in which the relative humidity that corresponds to a
given solution is guaranteed with the accuracy of several tenths of percent.
The analysis of the results of the laboratory examination of the sensors and
extended sets of measurements made with the Pulkovo cell made it possible to
conclude that in measurements in the interval of relative humidity 40-80%, the
~5% accuracy of the measurements for the water vapor content is reached.
Further paths are indicated for the increase of the accuracy of measurements
and extending the interval of the relative humidity, in which accurate
measurements may be carried out.
[4]
oai:arXiv.org:1010.3663 [pdf] - 243771
Combined Daily Monitoring of Aerosol Optical Depths and Water Vapour
Column Content during LACE 98 and LITFASS 98 Experiments
Submitted: 2010-10-18, last modified: 2010-10-19
During summer of 1998 two large-scale complex campaigns, LITFASS98 (May 25th
to June 22nd) and LACE98 (July 13th to August 14th), took place at the
Meteorological Observatory Lindenberg (MOL). The aim of both experiments focus
on the intensive daily observations of atmospheric conditions and the
determination of their fundamental meteorological parameters in the vertical
column over Lindenberg (Lindenberg's Column). About 20 German research
institutions and addition one from the Netherlands, Austria and Russia
participated at the experiments. A wide variety of ground-based instruments was
operated in Lindenberg and Falkenberg, including LIDARs, microwave radiometer
and radiosondes complemented by tethered balloons and aircraft measurements.
For the first time the star- and sunphotometer of MOL were used together with
other geophysical tools. The observations with both photometers were carried
out practically every day and night except during absolutely overcast
conditions. The observed data were processed immediately by a series of
programs developed at Pulkovo Observatory (Russia), and the results (daily
variations of aerosol optical depths and water vapour column content) were
presented at daily briefings. The comparison of these results with radiosonde
and microwave radiometer data demonstrated the usefulness of photometer data
for the calibration of other ground-based observations and satellite
measurements.
[5]
oai:arXiv.org:1010.4068 [pdf] - 245019
Monitoring of the terrestrial atmospheric characteristics with using of
stellar and solar photometry
Submitted: 2010-10-19
On the basis of experience acquired at creation of the Pulkovo
Spectrophotometric Catalog the method of investigation of a terrestrial
atmospheric components (aerosols and water vapor) in night time are designed.
For these purposes the small-sized photometers were created. Carried out in
1995-1999{\Gamma}.{\Gamma}. series of night and daily monitoring of the
atmospheric condition in Pulkovo, in MGO by A.I.Voejkov., in Germany (complex
experiments LITFASS 98 and LACE 98) confirmed suitability of devices,
techniques of observations and their reduction designed in Pulkovo Observatory
for the solution of geophysical and ecological problems. A final aim of this
work - creation of small-sized automatic complexes (telescope + photometer),
which would be rightful component of meteorological observatories. Such
complexes will work without the help of the observer and would provide the
daily monitoring of a terrestrial atmosphere.