oai:arXiv.org:1909.04769 [pdf] - 2025793
Photometric precision of a Si:As impurity band conduction mid-infrared
detector and application to transit spectroscopy
Submitted: 2019-09-10
Transit spectroscopy is the most promising path toward characterizing nearby
terrestrial planets at mid-infrared wavelengths in the next 20 years. The
Spitzer Space telescope has achieved moderately good mid-infrared photometric
precision in observations of transiting planets, but the intrinsic photometric
stability of mid-IR detectors themselves has not been reported in the
scientific or technical literature. Here, we evaluated the photometric
precision of a JWST MIRI prototype mid-infrared Si:As impurity band conduction
detector, using time-series data taken under flood illumination. These
measurements of photometric precision were conducted over periods of 10 hours,
representative of the time required to observe an exoplanet transit. After
selecting multiple sub-regions with a size of 10x10 pixels and compensating for
a gain change caused by our warm detector control electronics for the selected
sub-regions, we found that the photometric precision was limited to 26.3ppm at
high co-added signal levels due to a gain variation caused by our warm detector
control electronics. The photometric precision was improved up to 12.8ppm after
correcting for the gain drift. We also translated the photometric precision to
the expected spectro-photometric precision, assuming that an optimized
densified pupil spectrograph is used in transit observations. We found that the
spectro-photometric precision of an optimized densified pupil spectrograph when
used in transit observations is expected to be improved by the square root of
the number of pixels per a spectral resolution element. At the high co-added
signal levels, the total noise could be reduced down to 7ppm, which was larger
by a factor of 1.3 than the ideal performance that was limited by the Poisson
noise and readout noise. The systematic noise hidden behind the simulated
transit spectroscopy was 1.7ppm.
[2]
oai:arXiv.org:1811.11313 [pdf] - 1791684
SOFIA - HIRMES: Looking forward to the HIgh-Resolution Mid-infrarEd
Spectrometer
Richards, Samuel N.;
Moseley, Samuel H.;
Stacey, Gordon;
Greenhouse, Matthew;
Kutyrev, Alexander;
Arendt, Richard;
Atanasoff, Hristo;
Banks, Stuart;
Brekosky, Regis P.;
Brown, Ari-David;
Bulcha, Berhanu;
Cazeau, Tony;
Choi, Michael;
Colazo, Felipe;
Engler, Chuck;
Hadjimichael, Theodore;
Hays-Wehle, James;
Henderson, Chuck;
Hsieh, Wen-Ting;
Huang, Jeffrey;
Jenstrom, Iver;
Kellogg, Jim;
Kimball, Mark;
Kovacs, Attila;
Leiter, Steve;
Maher, Steve;
McMurray, Robert;
Melnick, Gary J.;
Mentzell, Eric;
Mikula, Vilem;
Miller, Timothy M.;
Nagler, Peter;
Nikola, Thomas;
Oxborrow, Joseph;
Pontoppidan, Klaus M.;
Rangwala, Naseem;
Rhodes, Alan;
Roberge, Aki;
Rosner, Stefan;
Rostem, Karwan;
Rustemeyer, Nancy;
Sharp, Elmer;
Sparr, Leroy;
Stevanovic, Dejan;
Taraschi, Peter;
Temi, Pasquale;
Vacca, William D.;
de Lorenzo, Jordi Vila Hernandez;
Wohler, Bill;
Wollack, Edward J.;
Wilks, Shannon
Submitted: 2018-11-27
The HIgh-Resolution Mid-infrarEd Spectrometer (HIRMES) is the 3rd Generation
Instrument for the Stratospheric Observatory For Infrared Astronomy (SOFIA),
currently in development at the NASA Goddard Space Flight Center (GSFC), and
due for commissioning in 2019. By combining direct-detection Transition Edge
Sensor (TES) bolometer arrays, grating-dispersive spectroscopy, and a host of
Fabry-Perot tunable filters, HIRMES will provide the ability for High
Resolution (R~100,000), Mid-Resolution (R~10,000), and Low-Resolution (R~600)
slit-spectroscopy, and 2D Spectral Imaging (R~2000 at selected wavelengths)
over the 25 - 122 {\mu}m mid-far infrared waveband. The driving science
application is the evolution of proto-planetary systems via measurements of
water-vapor, water-ice, deuterated hydrogen (HD), and neutral oxygen lines.
However, HIRMES has been designed to be as flexible as possible to cover a wide
range of science cases that fall within its phase-space, all whilst reaching
sensitivities and observing powers not yet seen thus far on SOFIA, providing
unique observing capabilities which will remain unmatched for decades.
[3]
oai:arXiv.org:1608.04323 [pdf] - 1456182
Proton irradiation results for long-wave HgCdTe infrared detector arrays
for NEOCam
Submitted: 2016-08-15
HgCdTe detector arrays with a cutoff wavelength of ~10 ${\mu}$m intended for
the NEOCam space mission were subjected to proton beam irradiation at the
University of California Davis Crocker Nuclear Laboratory. Three arrays were
tested - one with 800 $\mu$m substrate intact, one with 30 $\mu$m substrate,
and one completely substrate-removed. The CdZnTe substrate, on which the HgCdTe
detector is grown, has been shown to produce luminescence in shorter wave
HgCdTe arrays that causes elevated signal in non-hit pixels when subjected to
proton irradiation. This testing was conducted to ascertain whether or not full
substrate removal is necessary. At the dark level of the dewar, we detect no
luminescence in non-hit pixels during proton testing for both the
substrate-removed detector array and the array with 30 ${\mu}$m substrate. The
detector array with full 800 ${\mu}$m substrate exhibited substantial
photocurrent for a flux of 103 protons/cm$^2$-s at a beam energy of 18.1 MeV (~
750 e$^-$/s) and 34.4 MeV ($\sim$ 65 e$^-$/s). For the integrated space-like
ambient proton flux level measured by the Spitzer Space Telescope, the
luminescence would be well below the NEOCam dark current requirement of <200
e$^-$/s, but the pattern of luminescence could be problematic, possibly
complicating calibration.
[4]
oai:arXiv.org:astro-ph/0008271 [pdf] - 37587
Validation of Up-the-Ramp Sampling with Cosmic Ray Rejection on IR
Detectors
Offenberg, J. D.;
Fixsen, D. J.;
Rauscher, B. J.;
Forrest, W. J.;
Hanisch, R. J.;
Mather, J. C.;
McKelvey, M. E.;
McMurray, R. E.;
Nieto-Santisteban, M. A.;
Pipher, J. L.;
Sengupta, R.;
Stockman, H. S.
Submitted: 2000-08-17, last modified: 2000-10-25
We examine cosmic ray rejection methodology on data collected from InSb and
Si:As detectors. The application of an Up-the-Ramp sampling technique with
cosmic ray identification and mitigation is the focus of this study. This
technique is valuable for space-based observatories which are exposed to
high-radiation environments. We validate the Up-the-Ramp approach on
radiation-test data sets with InSb and Si:As detectors which were generated for
SIRTF. The Up-the-Ramp sampling method studied in this paper is over 99.9%
effective at removing cosmic rays and preserves the structure and photometric
quality of the image to well within the measurement error.