Normalized to: Lapolla, M.
[1]
oai:arXiv.org:1001.4644 [pdf] - 33160
Cryogenic environment and performance for testing the Planck radiometers
Terenzi, L.;
Lapolla, M.;
Laaninen, M.;
Battaglia, P.;
Cavaliere, F.;
De Rosa, A.;
Hughes, N.;
Jukkala, P.;
Kilpiä, V. -H.;
Morgante, G.;
Tomasi, M.;
Varis, J.;
Bersanelli, M.;
Butler, R. C.;
Ferrari, F.;
Franceschet, C.;
Leutenegger, P.;
Mandolesi, N.;
Mennella, A.;
Silvestri, R.;
Stringhetti, L.;
Tuovinen, J.;
Valenziano, L.;
Villa, F.
Submitted: 2010-01-26
This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst
The Planck LFI Radiometer Chain Assemblies (RCAs) have been calibrated in two
dedicated cryogenic facilities. In this paper the facilities and the related
instrumentation are described. The main satellite thermal interfaces for the
single chains have to be reproduced and stability requirements have to be
satisfied. Setup design, problems occurred and improving solutions implemented
are discussed. Performance of the cryogenic setup are reported.
[2]
oai:arXiv.org:1001.4646 [pdf] - 33161
Dynamic validation of the Planck/LFI thermal model
Tomasi, M.;
Cappellini, B.;
Gregorio, A.;
Colombo, F.;
Lapolla, M.;
Terenzi, L.;
Morgante, G.;
Bersanelli, M.;
Butler, R. C.;
Galeotta, S.;
Mandolesi, N.;
Maris, M.;
Mennella, A.;
Valenziano, L.;
Zacchei, A.
Submitted: 2010-01-26
The Low Frequency Instrument (LFI) is an array of cryogenically cooled
radiometers on board the Planck satellite, designed to measure the temperature
and polarization anisotropies of the cosmic microwave backgrond (CMB) at 30, 44
and 70 GHz. The thermal requirements of the LFI, and in particular the
stringent limits to acceptable thermal fluctuations in the 20 K focal plane,
are a critical element to achieve the instrument scientific performance.
Thermal tests were carried out as part of the on-ground calibration campaign at
various stages of instrument integration. In this paper we describe the results
and analysis of the tests on the LFI flight model (FM) performed at Thales
Laboratories in Milan (Italy) during 2006, with the purpose of experimentally
sampling the thermal transfer functions and consequently validating the
numerical thermal model describing the dynamic response of the LFI focal plane.
This model has been used extensively to assess the ability of LFI to achieve
its scientific goals: its validation is therefore extremely important in the
context of the Planck mission. Our analysis shows that the measured thermal
properties of the instrument show a thermal damping level better than
predicted, therefore further reducing the expected systematic effect induced in
the LFI maps. We then propose an explanation of the increased damping in terms
of non-ideal thermal contacts.
[3]
oai:arXiv.org:1001.4648 [pdf] - 33162
Planck-LFI radiometers tuning
Cuttaia, Francesco;
Mennella, Aniello;
Stringhetti, Luca;
Maris, Michele;
Terenzi, Luca;
Tomasi, Maurizio;
Villa, Fabrizio;
Bersanelli, Marco;
Butler, Christopher Reginald;
Cappellini, Benedetta;
Cuevas, Leticia Perez;
D'Arcangelo, Ocleto;
Davis, Richard;
Frailis, Marco;
Franceschet, Cristian;
Franceschi, Enrico;
Gregorio, Anna;
Hoyland, Roger;
Leonardi, Rodrigo;
Lowe, Stuart;
Mandolesi, Nazzareno;
Meinhold, Peter;
Mendes, Luis;
Roddis, Neil;
Sandri, Maura;
Valenziano, Luca;
Wilkinson, Althea;
Zacchei, Andrea;
Zonca, Andrea;
Battaglia, Paola;
De Nardo, Stefania;
Grassi, Stefano;
Lapolla, Marco;
Leutenegger, Paolo;
Miccolis, Maurizio;
Silvestri, Roberto
Submitted: 2010-01-26
"This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst"
This paper describes the Planck Low Frequency Instrument tuning activities
performed through the ground test campaigns, from Unit to Satellite Levels.
Tuning is key to achieve the best possible instrument performance and tuning
parameters strongly depend on thermal and electrical conditions. For this
reason tuning has been repeated several times during ground tests and it has
been repeated in flight before starting nominal operations. The paper discusses
the tuning philosophy, the activities and the obtained results, highlighting
developments and changes occurred during test campaigns. The paper concludes
with an overview of tuning performed during the satellite cryogenic test
campaign (Summer 2008) and of the plans for the just started in-flight
calibration.
[4]
oai:arXiv.org:1001.4653 [pdf] - 33163
Thermal susceptibility of the Planck-LFI receivers
Terenzi, L;
Salmon, M J;
Colin, A;
Mennella, A;
Morgante, G;
Tomasi, M;
Battaglia, P;
Lapolla, M;
Bersanelli, M;
Butler, R C;
Cuttaia, F;
D'Arcangelo, O;
Davis, R;
Franceschet, C;
Galeotta, S;
Gregorio, A;
Hughes, N;
Jukkala, P;
Kettle, D;
Laaninen, M;
Leutenegger, P;
Leonardi, R;
Mandolesi, N;
Maris, M;
Meinhold, P;
Miccolis, M;
Roddis, N;
Sambo, L;
Sandri, M;
Silvestri, R;
Tuovinen, J;
Valenziano, L;
Varis, J;
Villa, F;
Wilkinson, A;
Zonca, A
Submitted: 2010-01-26
This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst .
This paper describes the impact of the Planck Low Frequency Instrument front
end physical temperature fluctuations on the output signal. The origin of
thermal instabilities in the instrument are discussed, and an analytical model
of their propagation and impact on the receivers signal is described. The
experimental test setup dedicated to evaluate these effects during the
instrument ground calibration is reported together with data analysis methods.
Finally, main results obtained are discussed and compared to the requirements.
[5]
oai:arXiv.org:1001.4787 [pdf] - 33194
LFI Radiometric Chain Assembly (RCA) data handling "Rachel"
Malaspina, M.;
Franceschi, E.;
Battaglia, P.;
Binko, P.;
Butler, R. C.;
D'Arcangelo, O.;
Fogliani, S.;
Frailis, M.;
Franceschet, C.;
Galeotta, S.;
Gasparo, F.;
Gregorio, A.;
Lapolla, M.;
Leonardi, R.;
Maggio, G.;
Mandolesi, N.;
Manzato, P.;
Maris, M.;
Meharga, M.;
Meinhold, P.;
Morisset, N.;
Pasian, F.;
Perrotta, F.;
Rohlfs, R.;
Sandri, M.;
Tomasi, M.;
Turler, M.;
Zacchei, A.;
Zonca, A.
Submitted: 2010-01-26
This paper is part of the Prelaunch status LFI papers published on JINST
(http://www.iop.org/EJ/journal/-page=extra.proc5/1748-0221).
Planck's Low Frequency Instrument is an array of 22 pseudo-correlation
radiometers at 30, 44, and 70 GHz. Before integrating the overall array
assembly, a first set of tests has been performed for each radiometer chain
assembly (RCA), consisting of two radiometers. In this paper, we describe
Rachel, a software application which has been purposely developed and used
during the RCA test campaign to carry out both near-realtime on-line data
analysis and data storage (in FITS format) of the raw output from the
radiometric chains.