Normalized to: Kutsuma, H.
[1]
oai:arXiv.org:2005.08765 [pdf] - 2096547
A method to measure superconducting transition temperature of microwave
kinetic inductance detector by changing power of readout microwaves
Submitted: 2020-05-15
A microwave kinetic inductance detector (MKID) is a cutting-edge
superconducting detector, and its principle is based on a superconducting
resonator circuit. The superconducting transition temperature (Tc) of the MKID
is an important parameter because various MKID characterization parameters
depend on it. In this paper, we propose a method to measure the Tc of the MKID
by changing the applied power of the readout microwaves. A small fraction of
the readout power is deposited in the MKID, and the number of quasiparticles in
the MKID increases with this power. Furthermore, the quasiparticle lifetime
decreases with the number of quasiparticles. Therefore, we can measure the
relation between the quasiparticle lifetime and the detector response by
rapidly varying the readout power. From this relation, we estimate the
intrinsic quasiparticle lifetime. This lifetime is theoretically modeled by Tc,
the physical temperature of the MKID device, and other known parameters. We
obtain Tc by comparing the measured lifetime with that acquired using the
theoretical model. Using an MKID fabricated with aluminum, we demonstrate this
method at a 0.3 K operation. The results are consistent with those obtained by
Tc measured by monitoring the transmittance of the readout microwaves with the
variation in the device temperature. The method proposed in this paper is
applicable to other types, such as a hybrid-type MKID.
[2]
oai:arXiv.org:1907.03403 [pdf] - 1953635
A measurement method for responsivity of microwave kinetic inductance
detector by changing power of readout microwaves
Submitted: 2019-07-08
Superconducting detectors are a modern technology applied in various fields.
The microwave kinetic inductance detector (MKID) is one of cutting-edge
superconducting detector. It is based on the principle of a superconducting
resonator circuit. A radiation entering the MKID breaks the Cooper pairs in the
superconducting resonator, and the intensity of the radiation is detected as a
variation of the resonant condition. Therefore, calibration of the detector
responsivity, i.e., the variation of the resonant phase with respect to the
number of Cooper-pair breaks (quasiparticles), is important. We propose a
method for responsivity calibration. Microwaves used for the detector readout
locally raise the temperature in each resonator, which increases the number of
quasiparticles. Since the magnitude of the temperature rise depends on the
power of readout microwaves, the number of quasiparticles also depends on the
power of microwaves. By changing the power of the readout microwaves, we
simultaneously measure the phase difference and lifetime of quasiparticles. We
calculate the number of quasiparticles from the measured lifetime and by using
a theoretical formula. This measurement yields a relation between the phase
response as a function of the number of quasiparticles. We demonstrate this
responsivity calibration using the MKID maintained at 285mK. We also confirm
consistency between the results obtained using this method and conventional
calibration methods in terms of the accuracy.