oai:arXiv.org:1803.08702 [pdf] - 1674978
Fiber mode scrambler for the Subaru infrared Doppler instrument (IRD)
Submitted: 2018-03-23
We report the results of fiber mode scrambler experiments for the Infra-Red
Doppler instrument (IRD) on the Subaru 8.2-m telescope. IRD is a fiber-fed,
high precision radial velocity (RV) instrument to search for exoplanets around
nearby M dwarfs at near-infrared wavelengths. It is a high-resolution
spectrograph with an Echelle grating. The expected RV measurement precision is
1 m s-1 with a state of the art laser frequency comb for the wavelength
calibration. In IRD observations, one of the most significant instrumental
noise is a change of intensity distribution of multi-mode fiber exit, which
degrades RV measurement precision. To stabilize the intensity distribution of
fiber exit an introduction of fiber mode scrambler is mandatory. Several kinds
of mode scramblers have been suggested in previous research, though it is
necessary to determine the most appropriate mode scrambler system for IRD.
Thus, we conducted systematic measurements of performance for a variety of mode
scramblers, both static and dynamic. We tested various length multi-mode
fibers, an octagonal fiber, a double fiber scrambler, and two kinds of dynamic
scramblers, and their combinations. We report the performances of these mode
scramblers and propose candidate mode scrambler systems for IRD.