[1]  oai:arXiv.org:1603.02904  [pdf] - 1447692
Using 220Rn to calibrate liquid noble gas detectors
Comments: 6 pages,10 figures, to appear in Proceedings of the 14th international conference on Topics in Astroparticle and Underground Physics: TAUP2015, Torino, Italy
Submitted: 2016-03-07
In this paper, we describe 220Rn calibration source that was developed for liquid noble gas detectors. The key advantage of this source is that it can provide 212Bi-212Po consecutive events, which enables us to evaluate the vertex resolution of a detector at low energy by comparing low-energy events of 212Bi and corresponding higher-energy alpha-rays from 212Po. Since 220Rn is a noble gas, a hot metal getter can be used when introduced using xenon as the carrier gas. In addition, no long-life radioactive isotopes are left behind in the detector after the calibration is complete; this has clear advantage over the use of 222Rn which leaves long- life radioactivity, i.e., 210Pb. Using a small liquid xenon test chamber, we developed a system to introduce 220Rn via the xenon carrier gas; we demonstrated the successful introduction of 6 times 10^2 220Rn atoms in our test environment.