Normalized to: Horansky, R.
[1]
oai:arXiv.org:1310.7287 [pdf] - 1180273
High-resolution gamma-ray spectroscopy with a microwave-multiplexed
transition-edge sensor array
Noroozian, Omid;
Mates, John A. B.;
Bennett, Douglas A.;
Brevik, Justus A.;
Fowler, Joseph W.;
Gao, Jiansong;
Hilton, Gene C.;
Horansky, Robert D.;
Irwin, Kent D.;
Kang, Zhao;
Schmidt, Daniel R.;
Vale, Leila R.;
Ullom, Joel N.
Submitted: 2013-10-27
We demonstrate very high resolution photon spectroscopy with a
microwave-multiplexed two-pixel transition-edge sensor (TES) array. We measured
a $^{153}$Gd photon source and achieved an energy resolution of 63 eV
full-width-at-half-maximum at 97 keV and an equivalent readout system noise of
86 pA/$\sqrt{\text{Hz}}$ at the TES. The readout circuit consists of
superconducting microwave resonators coupled to radio-frequency
superconducting-quantum-interference-devices (SQUID) and transduces changes in
input current to changes in phase of a microwave signal. We use flux-ramp
modulation to linearize the response and evade low-frequency noise. This
demonstration establishes one path for the readout of cryogenic X-ray and
gamma-ray sensor arrays with more than $10^3$ elements and spectral resolving
powers $R=\lambda/\Delta\lambda > 10^3$.
[2]
oai:arXiv.org:1212.1738 [pdf] - 600827
Filters for High Rate Pulse Processing
Submitted: 2012-12-07
We introduce a filter-construction method for pulse processing that differs
in two respects from that in standard optimal filtering, in which the average
pulse shape and noise-power spectral density are combined to create a
convolution filter for estimating pulse heights. First, the proposed filters
are computed in the time domain, to avoid periodicity artifacts of the discrete
Fourier transform, and second, orthogonality constraints are imposed on the
filters, to reduce the filtering procedure's sensitivity to unknown baseline
height and pulse tails. We analyze the proposed filters, predicting energy
resolution under several scenarios, and apply the filters to high-rate pulse
data from gamma-rays measured by a transition-edge-sensor microcalorimeter.