First Commissioning of a Cryogenic Distillation Column for Low
Radioactivity Underground Argon
Back, H. O.;
Alexander, T.;
Alton, A.;
Condon, C.;
de Haas, E.;
Galbiati, C.;
Goretti, A.;
Hohmann, T.;
Ianni, An.;
Kendziora, C.;
Loer, B.;
Montanari, D.;
Mosteiro, P.;
Pordes, S.
Submitted: 2012-04-26, last modified: 2012-05-01
We report on the performance and commissioning of a cryogenic distillation
column for low radioactivity underground argon at Fermi National Accelerator
Laboratory. The distillation column is designed to accept a mixture of argon,
helium, and nitrogen and return pure argon with a nitrogen contamination less
than 10 ppm. In the first commissioning, we were able to run the distillation
column in a continuous mode and produce argon that is 99.9% pure. After running
in a batch mode, the argon purity was increased to 99.95%, with 500 ppm of
nitrogen remaining. The efficiency of collecting the argon from the gas mixture
was between 70% and 81%, at an argon production rate of 0.84-0.98 kg/day.