Normalized to: Heyne, B.
[1]
oai:arXiv.org:1805.01394 [pdf] - 1747907
100 GHz Room-Temperature Laboratory Emission Spectrometer
Submitted: 2018-05-03
We present first results of a new heterodyne spectrometer dedicated to
high-resolution spectroscopy of molecules of astrophysical importance. The
spectrometer, based on a roomtemperature heterodyne receiver, is sensitive to
frequencies between 75 and 110 GHz with an instantaneous bandwidth of currently
2.5 GHz in a single sideband. The system performance, in particular the
sensitivity and stability, is evaluated. Proof of concept of this spectrometer
is demonstrated by recording the emission spectrum of methyl cyanide, CH3CN.
Compared to state-of-the-art radio telescope receivers the instrument is less
sensitive by about one order of magnitude. Nevertheless, the capability for
absolute intensity measurements can be exploited in various experiments, in
particular for the interpretation of the ever richer spectra in the ALMA era.
The ease of operation at room-temperature allows for long time integration, the
fast response time for integration in chirped pulse instruments or for
recording time dependent signals. Future prospects as well as limitations of
the receiver for the spectroscopy of complex organic molecules (COMs) are
discussed.