Normalized to: Guenter, E.
[1]
oai:arXiv.org:1806.07361 [pdf] - 1703845
Comparing the emission spectra of U and Th hollow cathode lamps and a
new U line-list
Submitted: 2018-06-19, last modified: 2018-06-22
Thorium hollow cathode lamps (HCLs) are used as frequency calibrators for
many high resolution astronomical spectrographs, some of which aim for Doppler
precision at the 1 m/s level. We aim to determine the most suitable combination
of elements (Th or U, Ar or Ne) for wavelength calibration of astronomical
spectrographs, to characterize differences between similar HCLs, and to provide
a new U line-list. We record high resolution spectra of different HCLs using a
Fourier transform spectrograph: (i) U-Ne, U-Ar, Th-Ne, and Th-Ar lamps in the
spectral range from 500 to 1000 nm and U-Ne and U-Ar from 1000 to 1700 nm; (ii)
we systematically compare the number of emission lines and the line intensity
ratio for a set of 12 U-Ne HCLs; and (iii) we record a master spectrum of U-Ne
to create a new U line-list. Uranium lamps show more lines suitable for
calibration than Th lamps from 500 to 1000 nm. The filling gas of the lamps
significantly affects their performance because Ar and Ne lines contaminate
different spectral regions. We find differences (up to 88 %) in the line
intensity of U lines in different lamps from the same batch. We find 8239
isolated lines between 500 and 1700 nm that we attribute to U, 3379 of which
were not contained in earlier line-lists. The U line-list is available at the
http://www.astro.physik.uni-goettingen.de/research/U_atlas . We suggest using a
combination of U-Ne and U-Ar lamps to wavelength-calibrate astronomical
spectrographs up to 1000 nm. From 1000 to 1700 nm, U-Ne shows better
properties. The differences in line strength between different HCLs underline
the importance of characterizing HCLs in the laboratory. The new 3379 U lines
can significantly improve the radial velocity precision of astronomical
spectrographs.