Normalized to: Gräff, D.
[1]
oai:arXiv.org:1707.08923 [pdf] - 1586466
Complex spectral line profiles resulting from cryogenic deformation of
the SINFONI/SPIFFI diffraction gratings
Submitted: 2017-07-27
The integral field spectrograph, SPIFFI, has complex line profile shapes that
vary with wavelength and pixel scale, the origins of which have been sought
since the instrument construction. SPIFFI is currently operational as part of
SINFONI at the VLT, and will be upgraded and incorporated into the new VLT
instrument ERIS. We conducted an investigation of the line profiles based on
measurements we could take with the instrument calibration unit, as well as
laboratory measurements of spare SPIFFI optical components. Cryogenic
measurements of a spare SPIFFI diffraction grating showed significant periodic
deformation. These measurements match the cryogenic deformation expected from
bimetallic bending stress based on a finite element analysis of the
lightweighted grating blank. The periodic deformation of the grating surface
gives rise to satellite peaks in the diffraction pattern of the grating. An
optical simulation including the cryogenic grating deformation reproduces the
behavior of the SPIFFI line profiles with both wavelength and pixel scale as
measured with the instrument calibration unit. The conclusion is that cryogenic
deformation of the diffraction gratings is responsible for the non-ideal line
profiles, and that the diffraction gratings should be replaced during the
upgrade for optimal instrument performance.
[2]
oai:arXiv.org:1612.04680 [pdf] - 1533396
Upgrade and Characterization of the SPIFFI/SINFONI optics
Submitted: 2016-12-14
The SPIFFI integral field spectrometer is operated as a subunit of the AO
instrument SINFONI on the VLT. It will be upgraded and reused as SPIFFIER in
the new VLT instrument ERIS. In January 2016 an early upgrade of several
optical components was carried out to make use of new technological
developments before ERIS will be commissioned in 2020.
This Master's thesis focuses on the determination and detailed description of
the gain in performance due to the upgrade, with a particular focus on
measurements of the exchanged diamond turned mirrors of SPIFFI. An analysis of
the surface deformation of these mirrors is done, followed by a determination
of the resulting collimator wavefront and its influence on the line spread
function of the spectrometer. A careful analysis on the shapes, the variations
and the changes of the spectral line profiles due to the upgrade is done.
Furthermore, the spectral resolution, the image quality an the relative pre-
and post- upgrade throughput ratio are evaluated. Finally these observed
quantities are linked to critical optical components that limit the performance
of the instrument.
[3]
oai:arXiv.org:1608.02457 [pdf] - 1507611
Making SPIFFI SPIFFIER: Upgrade of the SPIFFI instrument for use in ERIS
and performance analysis from re-commissioning
George, E. M.;
Gräff, D.;
Feuchtgruber, H.;
Hartl, M.;
Eisenhauer, F.;
Buron, A.;
Davies, R.;
Genzel, R.;
Huber, H.;
Rau, C.;
Plattner, M.;
Wiezorrek, E.;
Weisz, H.;
Amico, P.;
Glindeman, A.;
Hau, G.;
Kuntschner, H.;
Modigliani, A.
Submitted: 2016-08-08, last modified: 2016-11-02
SPIFFI is an AO-fed integral field spectrograph operating as part of SINFONI
on the VLT, which will be upgraded and reused as SPIFFIER in the new VLT
instrument ERIS. In January 2016, we used new technology developments to
perform an early upgrade to optical subsystems in the SPIFFI instrument so
ongoing scientific programs can make use of enhanced performance before ERIS
arrives in 2020. We report on the upgraded components and the performance of
SPIFFI after the upgrade, including gains in throughput and spatial and
spectral resolution. We show results from re-commissioning, highlighting the
potential for scientific programs to use the capabilities of the upgraded
SPIFFI. Finally, we discuss the additional upgrades for SPIFFIER which will be
implemented before it is integrated into ERIS.