Normalized to: Ganjam, S.
[1]
oai:arXiv.org:1903.06517 [pdf] - 2060101
Single Electron-Hole Pair Sensitive Silicon Detector with Surface Event
Rejection
Hong, Ziqing;
Ren, Runze;
Kurinsky, Noah;
Figueroa-Feliciano, Enectali;
Wills, Lise;
Ganjam, Suhas;
Mahapatra, Rupak;
Mirabolfathi, Nader;
Nebolsky, Brian;
Pinckney, H. Douglas;
Platt, Mark
Submitted: 2019-03-15, last modified: 2020-02-27
We demonstrate single electron-hole pair resolution in a single-sided,
contact-free 1 cm$^2$ by 1 mm thick Si crystal operated at 48 mK, with a
baseline energy resolution of 3 eV. This crystal can be operated at voltages in
excess of $\pm50$ V, resulting in a measured charge resolution of 0.06
electron-hole pairs. The high aluminum coverage ($\sim$70%) of this device
allows for the discrimination of surface events and separation of events
occurring near the center of the detector from those near the edge. We use this
discrimination ability to show that non-quantized dark events seen in previous
detectors of a similar design are likely dominated by charge leakage along the
side wall of the device.
[2]
oai:arXiv.org:1805.10403 [pdf] - 1689861
A large-diameter cryogenic rotation stage for half-wave plate
polarization modulation on the POLARBEAR-2 experiment
Hill, Charles A.;
Kusaka, Akito;
Barton, Paul;
Bixler, Bryce;
Droster, Alex G.;
Flament, Mael;
Ganjam, Suhas;
Jadbabaie, Arian;
Jeong, Oliver;
Lee, Adrian T.;
Madurowicz, Alex;
Matsuda, Fred T.;
Matsumura, Tomotake;
Rutkowski, Adam;
Sakurai, Yuki;
Sponseller, Danielle R.;
Suzuki, Aritoki;
Tat, Raymond
Submitted: 2018-05-25
We describe the design of a cryogenic rotation stage (CRS) for use with the
cryogenic half-wave plate (CHWP) polarization modulator on the POLARBEAR-2b and
POLARBEAR-2c (PB2b/c) cosmic microwave background (CMB) experiments, the second
and third installments of the Simons Array. Rapid modulation of the CMB
polarization signal using a CHWP suppresses 1/f contamination due to
atmospheric turbulence and allows a single polarimeter to measure both
polarization states, mitigating systematic effects that arise when differencing
orthogonal detectors. To modulate the full detector array while avoiding excess
photon loading due to thermal emission, the CHWP must have a clear-aperture
diameter of > 450 mm and be cooled to < 100 K. We have designed a
454-mm-clear-aperture, < 65 K CRS using a superconducting magnetic bearing
driven by a synchronous magnetic motor. We present the specifications for the
CRS, its interfacing to the PB2b/c receiver cryostat, its performance in a
stand-alone test, and plans for future work.