oai:arXiv.org:1912.06571 [pdf] - 2014538
Large-area Si(Li) Detectors for X-ray Spectrometry and Particle Tracking
for the GAPS Experiment
Rogers, Field;
Xiao, Mengjiao;
Perez, Kerstin;
Boggs, Steven;
Erjavec, Tyler;
Fabris, Lorenzo;
Fuke, Hideyuki;
Hailey, Charles J.;
Kozai, Masayoshi;
Lowell, Alex;
Madden, Norman;
Manghisoni, Massimo;
McBride, Steve;
Re, Valerio;
Riceputi, Elisa;
Saffold, Nathan;
Shimizu, Yuki;
Zampa, Gianluigi
Submitted: 2019-12-13
Large-area lithium-drifted silicon (Si(Li)) detectors, operable 150{\deg}C
above liquid nitrogen temperature, have been developed for the General
Antiparticle Spectrometer (GAPS) balloon mission and will form the first such
system to operate in space. These 10 cm-diameter, 2.5 mm-thick multi-strip
detectors have been verified in the lab to provide <4 keV FWHM energy
resolution for X-rays as well as tracking capability for charged particles,
while operating in conditions (~-40{\deg}C and ~1 Pa) achievable on a
long-duration balloon mission with a large detector payload. These
characteristics enable the GAPS silicon tracker system to identify cosmic
antinuclei via a novel technique based on exotic atom formation, de-excitation,
and annihilation. Production and large-scale calibration of ~1000 detectors has
begun for the first GAPS flight, scheduled for late 2021. The detectors
developed for GAPS may also have other applications, for example in heavy
nuclei identification.
[2]
oai:arXiv.org:1906.00054 [pdf] - 1995412
Large-area Si(Li) detectors for X-ray spectrometry and particle tracking
in the GAPS experiment
Rogers, Field;
Xiao, Mengjiao;
Perez, Kerstin M.;
Boggs, Steven;
Erjavec, Tyler;
Fabris, Lorenzo;
Fuke, Hideyuki;
Hailey, Charles J.;
Kozai, Masayoshi;
Lowell, Alex;
Madden, Norman;
Manghisoni, Massimo;
McBride, Steve;
Re, Valerio;
Riceputi, Elisa;
Saffold, Nathan;
Shimizu, Yuki
Submitted: 2019-05-31, last modified: 2019-11-08
The first lithium-drifted silicon (Si(Li)) detectors to satisfy the unique
geometric, performance, and cost requirements of the General Antiparticle
Spectrometer (GAPS) experiment have been produced by Shimadzu Corporation. The
GAPS Si(Li) detectors will form the first large-area, relatively
high-temperature Si(Li) detector system with sensitivity to X-rays to operate
at high altitude. These 10 cm-diameter, 2.5 mm-thick, 4- or 8-strip detectors
provide the active area, X-ray absorption efficiency, energy resolution, and
particle tracking capability necessary for the GAPS exotic-atom particle
identification technique. In this paper, the detector performance is validated
on the bases of X-ray energy resolution and reconstruction of cosmic minimum
ionizing particle (MIP) signals. We use the established noise model for
semiconductor detectors to distinguish sources of noise due to the detector
from those due to signal processing electronics. We demonstrate that detectors
with either 4 strips or 8 strips can provide the required $\lesssim$4 keV
(FWHM) X-ray energy resolution at flight temperatures of $-35$ to
$-45^{\circ}$C, given the proper choice of signal processing electronics.
Approximately 1000 8-strip detectors will be used for the first GAPS Antarctic
balloon flight, scheduled for late 2021.
[3]
oai:arXiv.org:1906.05577 [pdf] - 1983873
Developing a mass-production model of large-area Si(Li) detectors with
high operating temperatures
Kozai, M.;
Fuke, H.;
Yamada, M.;
Perez, K.;
Erjavec, T.;
Hailey, C. J.;
Madden, N.;
Rogers, F.;
Saffold, N.;
Seyler, D.;
Shimizu, Y.;
Tokuda, K.;
Shimizu, Y.;
Tokuda, K.;
Xiao, M.
Submitted: 2019-06-13, last modified: 2019-08-27
This study presents a fabrication process for lithium-drifted silicon
(Si(Li)) detectors that, compared to previous methods, allows for mass
production at a higher yield, while providing a large sensitive area and low
leakage currents at relatively high temperatures. This design, developed for
the unique requirements of the General Antiparticle Spectrometer (GAPS)
experiment, has an overall diameter of 10 cm, with ~9 cm of active area
segmented into 8 readout strips, and an overall thickness of 2.5 mm, with
$\gtrsim$2.2 mm ($\gtrsim$90%) sensitive thickness. An energy resolution
$\lesssim$4 keV full-width at half-maximum (FWHM) for 20-100 keV X-rays is
required at the operating temperature ~-40C, which is far above the liquid
nitrogen temperatures conventionally used to achieve fine energy resolution.
High-yield production is also required for GAPS, which consists of
$\gtrsim$1000 detectors. Our specially-developed Si crystal and custom methods
of Li evaporation, diffusion and drifting allow for a thick, large-area and
uniform sensitive layer. We find that retaining a thin undrifted layer on the
p-side of the detector drastically reduces the leakage current, which is a
dominant component of the energy resolution at these temperatures. A guard-ring
structure and optimal etching of the detector surface are also confirmed to
suppress the leakage current. We report on the mass production of these
detectors that is ongoing now, and demonstrate it is capable of delivering a
high yield of ~90%.
[4]
oai:arXiv.org:1812.07255 [pdf] - 1800114
Development of Large-area Lithium-drifted Silicon Detectors for the GAPS
Experiment
Kozai, M.;
Fuke, H.;
Yamada, M.;
Erjavec, T.;
Hailey, C. J.;
Kato, C.;
Madden, N.;
Munakata, K.;
Perez, K.;
Rogers, F.;
Saffold, N.;
Shimizu, Y.;
Tokuda, K.;
Xiao, M.
Submitted: 2018-12-18
We have developed large-area lithium-drifted silicon (Si(Li)) detectors to
meet the unique requirements of the General Antiparticle Spectrometer (GAPS)
experiment. GAPS is an Antarctic balloon-borne mission scheduled for the first
flight in late 2020. The GAPS experiment aims to survey low-energy cosmic-ray
antinuclei, particularly antideuterons, which are recognized as essentially
background-free signals from dark matter annihilation or decay. The GAPS Si(Li)
detector design is a thickness of 2.5 mm, diameter of 10 cm and 8 readout
strips. The energy resolution of <4 keV (FWHM) for 20 to 100 keV X-rays at
temperature of -35 to -45 C, far above the liquid nitrogen temperatures
frequently used to achieve fine energy resolution, is required. We developed a
high-quality Si crystal and Li-evaporation, diffusion and drift methods to form
a uniform Li-drifted layer. Guard ring structure and optimal etching of the
surface are confirmed to suppress the leakage current, which is a main source
of noise. We found a thin un-drifted layer retained on the p-side effectively
suppresses the leakage current. By these developments, we succeeded in
developing the GAPS Si(Li) detector. As the ultimate GAPS instrument will
require >1000 10-cm diameter Si(Li) detectors to achieve high sensitivity to
rare antideuteron events, high-yield production is also a key factor for the
success of the GAPS mission.
[5]
oai:arXiv.org:1807.07912 [pdf] - 1719637
Fabrication of low-cost, large-area prototype Si(Li) detectors for the
GAPS experiment
Perez, Kerstin;
Aramaki, Tsuguo;
Hailey, Charles J.;
Carr, Rachel;
Erjavec, Tyler;
Fuke, Hideyuki;
Garvin, Amani;
Harper, Cassia;
Kewley, Glenn;
Madden, Norman;
Mechbal, Sarah;
Rogers, Field;
Saffold, Nathan;
Tajiri, Gordon;
Tokuda, Katsuhiko;
Williams, Jason;
Yamada, Minoru
Submitted: 2018-07-20
A Si(Li) detector fabrication procedure has been developed with the aim of
satisfying the unique requirements of the GAPS (General Antiparticle
Spectrometer) experiment. Si(Li) detectors are particularly well-suited to the
GAPS detection scheme, in which several planes of detectors act as the target
to slow and capture an incoming antiparticle into an exotic atom, as well as
the spectrometer and tracker to measure the resulting decay X-rays and
annihilation products. These detectors must provide the absorption depth,
energy resolution, tracking efficiency, and active area necessary for this
technique, all within the significant temperature, power, and cost constraints
of an Antarctic long-duration balloon flight. We report here on the fabrication
and performance of prototype 2"-diameter, 1-1.25 mm-thick, single-strip Si(Li)
detectors that provide the necessary X-ray energy resolution of $\sim$4 keV for
a cost per unit area that is far below that of previously-acquired commercial
detectors. This fabrication procedure is currently being optimized for the
4"-diameter, 2.5 mm-thick, multi-strip geometry that will be used for the GAPS
flight detectors.