Normalized to: Dorner, B.
[1]
oai:arXiv.org:1606.05640 [pdf] - 1486884
A model-based approach to the spatial and spectral calibration of
NIRSpec onboard JWST
Dorner, Bernhard;
Giardino, Giovanna;
Ferruit, Pierre;
de Oliveira, Catarina Alves;
Birkmann, Stephan M.;
Böker, Torsten;
De Marchi, Guido;
Gnata, Xavier;
Köhler, Jess;
Sirianni, Marco;
Jakobsen, Peter
Submitted: 2016-06-17
Context: The NIRSpec instrument for the James Webb Space Telescope (JWST) can
be operated in multiobject (MOS), long-slit, and integral field (IFU) mode with
spectral resolutions from 100 to 2700. Its MOS mode uses about a quarter of a
million individually addressable minislits for object selection, covering a
field of view of $\sim$9 $\mathrm{arcmin}^2$. Aims: The pipeline used to
extract wavelength-calibrated spectra from NIRSpec detector images relies
heavily on a model of NIRSpec optical geometry. We demonstrate how dedicated
calibration data from a small subset of NIRSpec modes and apertures can be used
to optimize this parametric model to the necessary levels of fidelity. Methods:
Following an iterative procedure, the initial fiducial values of the model
parameters are manually adjusted and then automatically optimized, so that the
model predicted location of the images and spectral lines from the fixed slits,
the IFU, and a small subset of the MOS apertures matches their measured
location in the main optical planes of the instrument. Results: The NIRSpec
parametric model is able to reproduce the spatial and spectral position of the
input spectra with high fidelity. The intrinsic accuracy (1-sigma, RMS) of the
model, as measured from the extracted calibration spectra, is better than 1/10
of a pixel along the spatial direction and better than 1/20 of a resolution
element in the spectral direction for all of the grating-based spectral modes.
This is fully consistent with the corresponding allocation in the spatial and
spectral calibration budgets of NIRSpec.
[2]
oai:arXiv.org:0808.0137 [pdf] - 15059
PYRAMIR: Calibration and operation of a pyramid near-infrared wavefront
sensor
Submitted: 2008-08-01
The concept of pyramid wavefront sensors (PWFS) has been around about a
decade by now. However, there is still a great lack of characterizing
measurements that allow the best operation of such a system under real life
conditions at an astronomical telescope. In this article we, therefore,
investigate the behavior and robustness of the pyramid infrared wavefront
sensor PYRAMIR mounted at the 3.5 m telescope at the Calar Alto Observatory
under the influence of different error sources both intrinsic to the sensor,
and arising in the preceding optical system. The intrinsic errors include
diffraction effects on the pyramid edges and detector read out noise. The
external imperfections consist of a Gaussian profile in the intensity
distribution in the pupil plane during calibration, the effect of an optically
resolved reference source, and noncommon-path aberrations. We investigated the
effect of three differently sized reference sources on the calibration of the
PWFS. For the noncommon-path aberrations the quality of the response of the
system is quantified in terms of modal cross talk and aliasing. We investigate
the special behavior of the system regarding tip-tilt control. From our
measurements we derive the method to optimize the calibration procedure and the
setup of a PWFS adaptive optics (AO) system. We also calculate the total
wavefront error arising from aliasing, modal cross talk, measurement error, and
fitting error in order to optimize the number of calibrated modes for on-sky
operations. These measurements result in a prediction of on-sky performance for
various conditions.