Normalized to: Dawson, C.
[1]
oai:arXiv.org:2001.02763 [pdf] - 2028471
An Impedance-Modulated Code-Division Microwave SQUID Multiplexer
Submitted: 2020-01-06
Large arrays of cryogenic detectors, including transition-edge sensors (TESs)
or magnetic micro-calorimeters (MMCs), are needed for future experiments across
a wide range of applications. Complexities in integration and cryogenic wiring
have driven efforts to develop cryogenic readout technologies with large
multiplexing factors while maintaining minimal readout noise. One such example
is the microwave SQUID multiplexer ($\mu$mux), which couples an incoming TES or
magnetic calorimeter signal to a unique GHz-frequency resonance that is
modulated in frequency. Here, we present a hybrid scheme combining the
microwave SQUID multiplexer with code division multiplexing: the
impedance-modulated code-division multiplexer (Z-CDM), which may enable an
order of magnitude increase in multiplexing factor particularly for
low-bandwidth signal applications.
[2]
oai:arXiv.org:1906.08814 [pdf] - 1904201
Exclusion Limits on Hidden-Photon Dark Matter near 2 neV from a
Fixed-Frequency Superconducting Lumped-Element Resonator
Phipps, A.;
Kuenstner, S. E.;
Chaudhuri, S.;
Dawson, C. S.;
Young, B. A.;
FitzGerald, C. T.;
Froland, H.;
Wells, K.;
Li, D.;
Cho, H. M.;
Rajendran, S.;
Graham, P. W.;
Irwin, K. D.
Submitted: 2019-06-20
We present the design and performance of a simple fixed-frequency
superconducting lumped-element resonator developed for axion and hidden photon
dark matter detection. A rectangular NbTi inductor was coupled to a Nb-coated
sapphire capacitor and immersed in liquid helium within a superconducting
shield. The resonator was transformer-coupled to a DC SQUID for readout. We
measured a quality factor of $\sim$40,000 at the resonant frequency of 492.027
kHz and set a simple exclusion limit on $\sim$2 neV hidden photons with kinetic
mixing angle $\varepsilon\gtrsim1.5\times10^{-9}$ based on 5.14 hours of
integrated noise. This test device informs the development of the Dark Matter
Radio, a tunable superconducting lumped-element resonator which will search for
axions and hidden photons over the 100 Hz to 300 MHz frequency range.
[3]
oai:arXiv.org:1610.09344 [pdf] - 1504575
Design Overview of the DM Radio Pathfinder Experiment
Silva-Feaver, Maximiliano;
Chaudhuri, Saptarshi;
Cho, Hsiao-Mei;
Dawson, Carl;
Graham, Peter;
Irwin, Kent;
Kuenstner, Stephen;
Li, Dale;
Mardon, Jeremy;
Moseley, Harvey;
Mule, Richard;
Phipps, Arran;
Rajendran, Surjeet;
Steffen, Zach;
Young, Betty
Submitted: 2016-10-28
We introduce the DM Radio, a dual search for axion and hidden photon dark
matter using a tunable superconducting lumped-element resonator. We discuss the
prototype DM Radio Pathfinder experiment, which will probe hidden photons in
the 500 peV (100 kHz)-50 neV (10 MHz) mass range. We detail the design of the
various components: the LC resonant detector, the resonant frequency tuning
procedure, the differential SQUID readout circuit, the shielding, and the
cryogenic mounting structure. We present the current status of the pathfinder
experiment and illustrate its potential science reach in the context of the
larger experimental program.