Normalized to: Bourhill, J.
[1]
oai:arXiv.org:1703.02179 [pdf] - 1581784
Sensitivity Characterisation of a Parametric Transducer for
Gravitational Wave Detection Through Optical Spring Effect
Submitted: 2017-03-06, last modified: 2017-07-24
We present the characterisation of the most recent parametric transducers
designed to enhance the Mario Schenberg Gravitational Wave Detector
sensitivity. The transducer is composed of a microwave re-entrant cavity that
attaches to the gravitational wave antenna via a rigid spring. It functions as
a three-mode mass-spring system; motion of the spherical antenna couples to a
50 $\mu m$ thick membrane, which converts its mechanical motion into a
frequency shift of the cavity resonance. Through the optical spring effect, the
microwave transducer frequency-displacement sensitivity was measured to be 726
$MHz/\mu$m at 4 K. The spherical antenna detection sensitivity is determined
analytically using the transducer amplification gain and equivalent
displacement noise in the test setup, which are 5.5 $\times$ 10$^{11} V/m$ and
$1.8 \times 10^{-19} m\sqrt{Hz}^{-1}$, respectively.