Barusso, Lorenzo Ferrari
Normalized to: Barusso, L.
1 article(s) in total. 13 co-authors. Median position in authors list is 10,0.
[1]
oai:arXiv.org:1912.06390 [pdf] - 2014523
The Demonstration Model of the ATHENA X-IFU Cryogenic AntiCoincidence
Detector
D'Andrea, Matteo;
Macculi, Claudio;
Torrioli, Guido;
Argan, Andrea;
Brienza, Daniele;
Lotti, Simone;
Minervini, Gabriele;
Piro, Luigi;
Biasotti, Michele;
Barusso, Lorenzo Ferrari;
Gatti, Flavio;
Rigano, Manuela;
Volpe, Angela;
Battistelli, Elia Stefano
Submitted: 2019-12-13
The Cryogenic AntiCoincidence Detector (CryoAC) of ATHENA X-IFU is designed
to reduce the particle background of the instrument and to enable the mission
science goals. It is a 4 pixel silicon microcalorimeter sensed by an Ir/Au TES
network. We have developed the CryoAC Demonstration Model, a prototype aimed to
probe the critical technologies of the detector, i.e. the suspended absorber
with an active area of 1 cm2; the low energy threshold of 20 keV; and the
operation connected to a 50 mK thermal bath with a power dissipation less than
40 nW. Here we report the test performed on the first CryoAC DM sample (namely
the AC-S10 prototype), showing that it is fully compliant with its
requirements.