oai:arXiv.org:1202.2628 [pdf] - 552768
Characterization of the Hamamatsu R11410-10 3-Inch Photomultiplier Tube
for Liquid Xenon Dark Matter Direct Detection Experiments
Lung, K.;
Arisaka, K.;
Bargetzi, A.;
Beltrame, P.;
Cahill, A.;
Genma, T.;
Ghag, C.;
Gordon, D.;
Sainz, J.;
Teymourian, A.;
Yoshizawa, Y.
Submitted: 2012-02-12, last modified: 2012-08-23
To satisfy the requirements of the next generation of dark matter detectors
based on the dual phase TPC, Hamamatsu, in close collaboration with UCLA, has
developed the R11410-10 photomultipler tube. In this work, we present the
detailed tests performed on this device. High QE (>30%) accompanied by a low
dark count rate (50 Hz at 0.3 PE) and high gain (10^7) with good single PE
resolution have been observed. A comprehensive screening measurement campaign
is ongoing while the manufacturer quotes a radioactivity of 20 mBq/PMT. These
characteristics show the R11410-10 to be particularly suitable for the
forthcoming zero background liquid xenon detectors.