Arslanyan, V.
Normalized to: Arslanyan, V.
1 article(s) in total. 14 co-authors. Median position in authors list is 14,0.
[1]
oai:arXiv.org:1203.5075 [pdf] - 1117487
FIRST, a fibered aperture masking instrument. I. First on-sky test
results
Huby, E.;
Perrin, G.;
Marchis, F.;
Lacour, S.;
Kotani, T.;
DuchĂȘne, G.;
Choquet, E.;
Gates, E. L.;
Woillez, J. M.;
Lai, O.;
Fédou, P.;
Collin, C.;
Chapron, F.;
Arslanyan, V.;
Burns, K. J.
Submitted: 2012-03-22
In this paper we present the first on-sky results with the fibered aperture
masking instrument FIRST. Its principle relies on the combination of spatial
filtering and aperture masking using single-mode fibers, a novel technique that
is aimed at high dynamic range imaging with high angular resolution. The
prototype has been tested with the Shane 3-m telescope at Lick Observatory. The
entrance pupil is divided into subpupils feeding single-mode fibers. The flux
injection into the fibers is optimized by a segmented mirror. The beams are
spectrally dispersed and recombined in a non-redundant exit configuration in
order to retrieve all contrasts and phases independently. The instrument works
at visible wavelengths between 600 nm and 760 nm and currently uses nine of the
30 43 cm subapertures constituting the full pupil. First fringes were obtained
on Vega and Deneb. Stable closure phases were measured with standard deviations
on the order of 1 degree. Closure phase precision can be further improved by
addressing some of the remaining sources of systematic errors. While the number
of fibers used in the experiment was too small to reliably estimate visibility
amplitudes, we have measured closure amplitudes with a precision of 10 % in the
best case. These first promising results obtained under real observing
conditions validate the concept of the fibered aperture masking instrument and
open the way for a new type of ground-based instrument working in the visible.
The next steps of the development will be to improve the stability and the
sensitivity of the instrument in order to achieve more accurate closure phase
and visibility measurements, and to increase the number of sub-pupils to reach
full pupil coverage.