Dust Impact Monitor (SESAME-DIM) on board Rosetta/Philae: Millimetric
particle flux at comet 67P/Churyumov-Gerasimenko
Hirn, Attila;
Albin, Thomas;
Apáthy, István;
Della Corte, Vincenzo;
Fischer, Hans-Herbert;
Flandes, Alberto;
Loose, Alexander;
Péter, Attila;
Seidensticker, Klaus J.;
Krüger, Harald
Submitted: 2016-05-20
The Philae lander of the Rosetta mission, aimed at the in situ investigation
of comet 67P/C-G, was deployed to the surface of the comet nucleus on 12 Nov
2014 at 2.99 AU heliocentric distance. The Dust Impact Monitor (DIM) as part of
the Surface Electric Sounding and Acoustic Monitoring Experiment (SESAME) on
the lander employed piezoelectric detectors to detect the submillimetre- and
millimetre-sized dust and ice particles emitted from the nucleus. We determine
the upper limit of the ambient flux of particles in the measurement range of
DIM based on the measurements performed with the instrument during Philae's
descent to its nominal landing site Agilkia at distances of about 22 km, 18 km,
and 5 km from the nucleus barycentre and at the final landing site Abydos. The
geometric factor of the DIM sensor is calculated assuming an isotropic ambient
flux of the submillimetre- and millimetre-sized particles. For the measurement
intervals when no particles were detected the maximum true impact rate was
calculated by assuming Poisson distribution of the impacts, and it was given as
the detection limit at a 95% confidence level. Based on measurements performed
with DIM, the upper limit of the flux of particles in the measurement range of
the instrument was of the order of $10^{-8}-10^{-7}m^{-2}s^{-1}sr^{-1}$ during
descent and $1.6\cdot10^{-9}m^{-2}s^{-1}sr^{-1}$ at Abydos on 13 and 14 Nov
2014. Considering particle speeds below escape velocity, the upper limit for
the volume density of particles in the measurement range of DIM was constrained
to $10^{-11}m^{-3}-10^{-12}m^{-3}$. Results of the calculations performed with
the GIPSI tool on the expected particle fluxes during the descent of Philae
were compatible with the non-detection of compact particles by the DIM
instrument.