Normalized to: Amorese, A.
[1]
oai:arXiv.org:1410.1587 [pdf] - 877008
Enhancing spatial resolution of soft x-ray CCD detectors by
single-photon centroid determination
Submitted: 2014-10-06
In Charge Coupled Device (CCD) detectors the electrons excited upon
absorption of a single x-ray photon quickly diffuse and generate charge-spots
often larger than pixel dimensions. In the soft x-ray range this phenomenon
drastically limits the effective spatial resolution to approximately 25\mu m,
irrespective of the pixel size. For very low fluence the charge-cloud centroid
determination can be used, on each individual spot, to estimate the actual
photon impact position with sub-pixel precision. The readout noise and speed,
together with the charge and spatial undersampling, are the main factors
limiting the accuracy of this procedure in commercial devices. We have
developed and extensively tested an algorithm for efficient centroid
reconstruction on images acquired by low noise detectors not designed for
single photon counting. We have thus measured a position uncertainty of 6-7\mu
m in CCDs with 13.5\mu m and 20.0\mu m pixel size, around 1 keV photon energy.
We have analyzed the centroid calculation performances by modelling the spot
generation process. We show how the resolution is affected by both random
uncertainty, mainly ascribed to the readout noise, and systematic error, due to
the undersampling of the charge spot. This study was motivated by the growing
need of of high-resolution high-sensitivity 2D position sensitive detectors in
resonant inelastic (soft) x-ray scattering (RIXS), an emergent synchrotron
radiation based spectroscopy.